![Huvitz HDR-9000 Service Manual Download Page 13](http://html1.mh-extra.com/html/huvitz/hdr-9000/hdr-9000_service-manual_2190280013.webp)
- 12 -
3.2.4 Checking Procedures for Data Communication
①
Checking with Ref/Keratometer.
- Check the Option menu of MRK or HRK for the communication.
Set the BPS option to 9600.
- Measure a model eye or a patient.
- Press PRINT button for transferring the measured data.
- Press RK button with SHIFT pressed on the Operation Panel after the MRK or HRK
finishes printing.
- Select and load the data received.
- Check if the data is correct.
②
Checking with Lensmeter.
- Check the Option menu of CLM or HLM for the communication.
Set the RS232 option.
Set the BPS option to 9600.
- Measure a lens.
- Press PRINT button for transferring the measured data.
- Press LM button with SHIFT pressed on the Operation Panel after the CLM or HLM
finishes printing.
- Select and load the data received.
- Check if the data is correct.
③
Checking with CCP-3100
- Check the System Settings menu of the Operation Panel for CCP-3100.
Set the Chart Connection option to SERIAL (it doesn't support IR option).
Set the Chart Device option to CCP-3100.
- Store the change by pressing the button OK or “HOME” button on the display.
- Press some chart buttons and check the response of CCP-3100.
④
Checking with CDC-4000
- Check the System Settings menu of the Operation Panel for CDC-4000.
Set the Chart Connection option to SERIAL (it doesn't support IR option).
Set the Chart Device option to CDC-4000.
- Store the change by pressing the button OK or “HOME” button on the display.
- Press some chart buttons and check the response of CDC-4000.
Summary of Contents for HDR-9000
Page 1: ...Service Manual Digital Refractor HDR 9000 HUVITZ Co Ltd...
Page 2: ...1...
Page 9: ...8 Engineer Mode Refer to appendix for more detailed information about this mode...
Page 17: ...16 Unscrew M2 6 screws and remove BACK CAP BACK CAP...
Page 18: ...17 Unscrew 3 M2 6 screws and remove TOP FRONT CASE TOP FRONT CASE...
Page 19: ...18 Unscrew 6 M2 6 screws and remove INNER CASE L INNER CASE R INNER CASE L INNER CASE R...
Page 30: ...29 4 6 Removing the LCD Panel Cover of the Operation Panel...
Page 32: ...31 4 8 Disassembling the Junction Box...
Page 37: ...36 5 1 2 Refractor Body Wiring Diagram...
Page 40: ...39 5 1 3 Junction Box Wiring Diagram...
Page 42: ...41...