10.30
Front Panel Connector Assembly refitting
1. Refitting is the reverse of removal. Take care not to damage the earthing
fingers. Renew cable ties.
10.31
Toco/Ultrasound Transducer Dismantling
1. Remove the 5 screws from rear of transducer (one under label). Invert case
assembly and lift top clear, care should be taken not to lift seal clear at this
stage.
10.32
Reassembly of Toco/Ultrasound Transducer
1. Ensure that seal is located in case before fitting transducer top and that seal
assembly is fitted correctly.
2. Fit transducer top and tighten screws to 20 cNm. Replace label.
F
Water resistance of US transducers can only be ensured if the units are
returned to Huntleigh Diagnostics for refurbishment.
10.33
Strain Gauge Assembly Removal
1. Remove case top as detailed in 10.31 and desolder grey cable from termination
PCB. Desolder strain gauge wires from PCB.
2. Remove 2 securing screws from strain gauge PCB and lift clear from assembly.
10.34
Strain Gauge Assembly Refitting
1. Fit strain gauge assembly noting orientation.
2. Solder wires to termination PCB and tighten screws to 40 cNm. Solder
connector cable to termination PCB. Align transducer as per 10.35.