Real time clock battery, bed number, address
g) Using , select desired option.
h) Using keys under LCD change setting.
i) Press menu button to escape or save as appropriate.
11.4
Print Head Test
A series of narrow stripes can be printed, this can be used as an aid to print head
alignment, this can be accessed as detailed in 8.3 steps a-f above.
11.5
Bed number
During normal operation the bed number is printed alongside the Huntleigh Diagnostics
footnote.
This can be changed as detailed in 8.3.
11.6
Real time clock battery
The real time clock battery can be disabled as detailed in 8.3.
This allows the life of the battery to be extended while in storage. The battery should
normally be turned on.
11.7
Address
The address of the monitor may be changed, however this is only used during factory
testing and must remain set to 10 for twins mode to function (BD4000 only).
Issue: 2 DRAFT
55
Rev: a
614365