2.3
FECG Testing
Apply 1500Vac between the;
a) chassis earth and leg electrode plate
b) instrument live terminal and FECG terminals
c) instrument neutral terminal and FECG terminals
The maximum allowable reading being 10
m
A over 60 secs.
If you require any assistance with safety testing your Huntleigh Diagnostics
equipment, contact Huntleigh Diagnostics. For the U.K. refer to the Health
Equipment Information Document No 95 – Code Of Practice For Acceptance Testing
Of Medical Equipment.
The following safety summary should be read before operating or carrying out any of
the procedures described in this manual:
4
Issue: 2 Draft
Rev: a
614365
Caution
Do not apply test voltages to main unit FECG input socket directly
Cautions
q
Do not use the
Baby Dopplex
in the presence of flammable gases such as
anaesthetic agents.
q
This product is
not
designed for sterile use. Do not use in the sterile field unless
additional barrier precautions are taken.
Do not
-
immerse main unit in any liquid. See section 2.4.
-
use solvent cleaners.
-
use high temperature sterilising processes (such as autoclaving).
-
use E-beam or gamma radiation sterilisation.