www.humancaregroup.com
25
ENGLISH
The metal sections of the frame can
be cleaned with a soft, damp cloth.
Cleaned surfaces should be wiped
dry each time. Do not use cleaning
products containing alcohol.
Check whether there is mechanical
damage to the wiring;
Check whether there is mechanical
damage to the housing of the control-
ler, battery and actuator;
Check whether the height adjustment
and leg adjustment buttons are
working properly.
All moving joints must be lubricated
once every six months or when they
start making a loud noise when used.
The moving joints include:
•
castors and actuators;
• bearing bushes and joints of the
legs and lift arm.
As a lubricant, we recommend using
commercially available penetrating
and lubricating products (e.g.
Wurth HHS 2000). Any spillage of
such lubricants must be removed
immediately with a dry cloth.
Threaded connections should be
inspected periodically – once every
six months – and, if necessary,
any accidental tolerance must
be remedied. Any inadvertent
tolerances must be reported to the
manufacturer's service center. The
product can no longer be used until
the cause of this tolerances has been
removed.
Maintenance of the frame
Maintenance of the frame
A service is required every 12 months or 8000
lifting cycles, whichever comes first and every
time after a fault/repair of the lift. Services,
including the electrical safety test, must be
carried out by a certified person.
Services should include as a minimum: