18
www.humancaregroup.com
seBastian II
Battery charge display and charging
Caution!
• The product cannot be used while charging.
•
During charging, the lifting function is
locked.
•
Do not use the lift for 1 hour after charging.
This will prolong the life of the battery.
•
The battery needs to be charged continuous
-
ly for at least 24 hour,
in the following situations:
a) Before using the lift for the first time.
b) Before being placed in storage (up to 3
months) without a power supply connec
-
tion.
c) Immediately after a storage period.
Battery charge indicator
The battery charge indicator has three status
indications. The battery indicators are either
yellow or green until switched off (2 minutes
after use).
Battery status 1:
The battery is full – no recharging required
(100-50%): the third indicator is green.
Battery status 2:
The battery is okay (50-25%) – recharging re
-
quired: the second indicator is yellow.
Battery status 3:
The battery is low (less than 25%) – recharging
required: the first indicator is yellow and a buzz
-
er sounds whenever a button is pressed.
Battery status 4:
The battery requires recharging. Some functions
of the lift are lost and it is only possible to lower
the lift arm. When in continued use, the indica
-
tor flashes and an audio signal is emitted.
The battery charge is indicated on the
hand control.
Battery level indicators on the hand
control