16
www.humancaregroup.com
seBastian II
Emergency stop
In the event of an emergency, immediately
press the emergency stop button.
Turn the button clockwise to unlock the emer
-
gency stop.
Electric emergency lowering function
If the button on the hand control for lowering
the lift arm does not work, you can use the
electric emergency lowering function. This can
be activated by using a pointed object (e.g. a
pen) to press the small button on the control
box. You will hear an audible click. Keep press-
ing until the required height has been reached.
Mechanical emergency lowering function
If the emergency lowering function via the
control box does not work, use the mechanical
lowering function. This will safely lower the
client in the event of a fault in the electrical sys-
tem. Lowering can be done manually by using
the lever located on the lower part of the main
cylinder. To operate, grasp the red part and pull
it up. Allow the lift arm to lower until the user
can be safely removed from the lift.
Caution!
After using the mechanical or electrical
emergency lowering function, you must
immediately contact your distributor
to schedule a visit by a service technician. The
product may only be used again after an autho
-
rized employee has carried out an inspection of
the product.
Emergency stop
ON CHARGE
Mechanical lower-
ing lever
Electrical emergency lowering