background image

2

II. Feature

Intelligent High-voltage insulation resistance tester, with 5 range: 500V, 1000V,

2500V, 5000V, 10000V, test up to 10TΩ.

● Strict accordance with the safety standards for design

Insulation resistance range 10TΩ @ 10 kv

● Short circuit current can be adjusted, maximum 5 mA (Up to 10mA short circuit current

product can be ordered.

● Automatic display polarization index (PI), Induced electric absorption ratio test values

(DAR), and can test the leakage current and capacitance.

● The anti-interference performance is superior. When the interference current reaches

2mA, the instrument still ensures the test accuracy.

Rapid discharge of capacitive test products, no need for manual discharge when testing

cables, automatic rapid discharge of instruments.

●2 power supply modes: the battery is powered by lithium ion battery, and the battery life

can reach 4 hours (10000V @100m test resistance). It can also be charged while in use.

In case of power failure, it can be automatically converted from AC power supply to

battery power supply

.

Insulation resistance simulation column.

Digital filtering functions, external influence deviation display value causing mitigation

point using filter function.

Perfect protection function, and the spare fuse prompt functions.

III. Specification Parameter

● Safety Standard

IEC 61010-1

CAT.IV 600V

IEC 61326

EMC specification: test, control and electrical equipment for testing

IEC60529 IP64

Outer closed state

● AC Power

220V±10%,50/60 HZ,20 VA

●battery

16.8V Lithium ion rechargeable battery

● Battery Life

10000V@100M

4 hour

Summary of Contents for HZJY-10K-I

Page 1: ...HZJY 10K I Insulation Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...Contents I Safe Use Items 1 II Feature 2 III Specification Parameter 2 IV Introduction Of Instrument 3 V Instrument Using 6 VI Packing List 7...

Page 3: ...operators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in...

Page 4: ...al discharge when testing cables automatic rapid discharge of instruments 2 power supply modes the battery is powered by lithium ion battery and the battery life can reach 4 hours 10000V 100m test res...

Page 5: ...range 20uF Capacitance testing accuracy 15 0 03uF Capacitor discharge rate from 10000V to 10V 0 5S F Insulation resistance testing range and accuracy T 23 5 C RH 45 75 range accuracy 500V 1000V 2500V...

Page 6: ...Power Off No operation 3 minutes automatic shutdown H V Set test voltage 500V 1000V 2500V 5000V 10000V TIME SetThe test time 00 30 01 00 10 00 30 00 Switching software keys to a new function or from...

Page 7: ...1 key the power button until into the contrast adjustment menu 3 Press 1 or 2 keys to adjust contrast until the right 4 Press the 4 key save and exit if you don t need to adjust the press 3 key to ex...

Page 8: ...oltage and start 3 The instrument automatically record 15S 60S and 600S resistance and calculate the polarization index after 600 seconds Polarization index calculation method PI R600 R60 test connect...

Page 9: ...7 VI Packing List No Item Qty 1 Main engine 1 2 Power line 1 3 Black test line 1 4 Red test line 1 5 Green test line 1...