3
● size
L x W x H
):
26cm x 20cm x 16cm
● weight
:
3kg
● Test voltage accuracy: the nominal value 100% to 110%
● Voltage measurement accuracy
:
5%+3V
● Current test range
:
10mA
● current measure precision
:
5%+0.2nA
● short-circuit current
:
5mA
● Capacitance test range
:
20uF
● Capacitance testing accuracy
:
15%+0.03uF
● Capacitor discharge rate
:
from 10000V to 10V
,
0.5S/µF
● Insulation resistance testing range and accuracy
(
T:23±5ºC
,
RH :45 – 75%
)
500V
1000V
2500V
5000V
10000V
Unspecified
<100k
<100k
<100k
<100k
<100k
5%
100k-10G 100k-20G 100k-100G 100k-100G
100k-200G
20%
10G -100G 20G-200G
100G-1T
100G-1T
200G-2T
Unspecified
> 100G
> 200G
> 1T
> 1T
> 2T
IV. Introduction Of Instrument
●
The Front Panel Description