background image

5

When the battery voltage shows just one more case should be timely charge, when

the battery displaying flashing is automatically cut off the power at any time, should

be recharged.

charge

AC socket provided meet the requirements of the AC power supply for panel,

you can recharge equipment charging, the power indicator, display the battery label,

in turn, from

(0) to (full) showed that when filled, displays (full)

Contrast Settings interface: LCD display gray level change due to temperature

changes, may affect the electronic contrast adjustment using the instrument Settings

Follow these steps to set up:

1. In shutdown state, hold key 2 (the 2 key).

2. Press the 1 key (the power button), until into the contrast adjustment menu.

3. Press 1 or 2 keys to adjust contrast, until the right .

4. Press the 4 key save and exit, if you don't need to adjust the press 3 key to exit.

Soft key

Description

Move the cursor, also used to increase or decrease in contrast

Move the cursor, also used to increase or decrease in contrast

Back

Return to the main menu

OK

Confirm to enter the current menu function or to determine the
current Settings and exit

Help menu system

HELP

Record the serial number of product, production date, version information.

Insulation testing interface

Digital

Insulation value

( >

greater than

; <

less than

)

Measured time

actual voltage

H.V

warning

Leakage curren

t

1mA=1000uA
1uA =1000nA

DAR

Capacitance

PI

Simulation column

When the test time or manual stopped, display shows as shown in the table of the soft

keys.

Summary of Contents for HZJY-10K-I

Page 1: ...HZJY 10K I Insulation Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...Contents I Safe Use Items 1 II Feature 2 III Specification Parameter 2 IV Introduction Of Instrument 3 V Instrument Using 6 VI Packing List 7...

Page 3: ...operators in instrument surface operation Please don t charged the connection test line low Low in the measurement or do not touch the circuit being measured immediately after the test may result in...

Page 4: ...al discharge when testing cables automatic rapid discharge of instruments 2 power supply modes the battery is powered by lithium ion battery and the battery life can reach 4 hours 10000V 100m test res...

Page 5: ...range 20uF Capacitance testing accuracy 15 0 03uF Capacitor discharge rate from 10000V to 10V 0 5S F Insulation resistance testing range and accuracy T 23 5 C RH 45 75 range accuracy 500V 1000V 2500V...

Page 6: ...Power Off No operation 3 minutes automatic shutdown H V Set test voltage 500V 1000V 2500V 5000V 10000V TIME SetThe test time 00 30 01 00 10 00 30 00 Switching software keys to a new function or from...

Page 7: ...1 key the power button until into the contrast adjustment menu 3 Press 1 or 2 keys to adjust contrast until the right 4 Press the 4 key save and exit if you don t need to adjust the press 3 key to ex...

Page 8: ...oltage and start 3 The instrument automatically record 15S 60S and 600S resistance and calculate the polarization index after 600 seconds Polarization index calculation method PI R600 R60 test connect...

Page 9: ...7 VI Packing List No Item Qty 1 Main engine 1 2 Power line 1 3 Black test line 1 4 Red test line 1 5 Green test line 1...