g.
PERFORMANCE TESTS
11. Uoe the following procedure (steps a through kb to perform a 1 PORT FULL CALI-
BRATION for the adapter on the TEST PORT connector of the HP 41952A (test
equipment).
a. Sequentially press the HP 4195A's CAL key, and the 'CORRECTN ors off', 'more
1/2', and 'CAL STD modify' softkeys,
b. Sequentially press the following keys to change the HP 4195A's stored standard,
calibration data,
'OPEN CAL STD', 0 , green shlftkey„ (comma), 1 , 0 , 8
EEX, - (minus), 1 , 5 , ENTER/EXECUTE
'SHORT CAL STD', 0 , green shiftkey„ (comma), 0 , ENTER/EXECUTE
c. Press the MENU key (once) and the 'TYPE tin log' softkey (twice).
d. Sequentially press the HP 4195A's CAL key, and the, 'S-PRMTR CAL menu' and
the 'ONE PORT FULL CAL' softkeys.
e. Connect the HP 850326's N(f) OPEN standard to the adapter on the TEST PORT
connector of the HP 41962A (test equipment),
f, P r e s s the 'OPEN' sollkey and the ENTER/EXECUTE key. Wait until the single
sweep.measurement is completed,
Connect the HP 850326's N(f) SHORT standard to the adapter on the TEST
PORT 'connector of the HP 41952A (test equipment).
Press the 'SHORT' soflkey and the ENTER/EXECUTE key. Walt until, the single
v‘ieep measurement is completed.
i. C o n n e c t tnie HP 850326's NW LOAD standard to the adapter on the TEST PORT
connector of the HP 41952A (test equipment).
Press the 'LOAD' softkey and the ENTER/EXECUTE key, Wait until the single
sweep measurement is completed.
k. Press the CAL key and the 'CORRECTN on oft' softkey.
12. Perform the following procedure (steps a through h) to test the Port Match for the RF
INPUT connector of the HP 41952A/B (under test).
4.17