I i
PERFORMANCE TESTS
I I. Press the 'CAL' key and the 'CORRECTN on off' solthey. "on" I i this softkey will
l' change to intensiflad green.
/
,
1
1,2,) Press the PROGRAM key, the 'EDIT' softkey and the UNTER/17XECUTE key, In se-
qudnCe. The HP 4195A'a display will change to the user progrep editor display,
,13. Enter the followl,ig HP 4195A program to have It display the Effective Source Match
test results. I i you need more Information on how to enter a HP 4195A program,
refer to the HP 4195A's Operation Manual, paragraph 6-4, USER PROGRAM (Auto
Sequence Program:ASP).
10 <RA,RB> = <MFOR,MFOI> • <TFSR,TFS$>
20 <RA,RB> = <RA,RB> <MFSR,MFSI> ‹TFOR,TFOI>
30 <RA,RB> a <RA,RB> + <MFLR,MFLI> • (<TFOR,TFOI> <TFSR,TESI>)
40 <RC,RD> = <MFOR,MFOI> • ‹TFOR,TFOI> • ‹TFSR,TFSI>
50 <PC,RD> = <RC,RD> <MFSR,MFSI> • <TFSR,TFSI> . <TFOR,TFOI>
60 <RE,RF> <RA,RB> / <RC,RD>
70 A = - 20 • LOG ( SOP,(RE • RE + RF • RF ) )
60 E N D
Note; T o display " < ", press the green shiftkey and < key.
To display " > ", press the green shiftkey and > key.
To display " , '', press the
green shiftkey and
, (comma) key.
To display " . '', press
the MATH OPERATOR
key and • . softkey,
To display " " , press the MATH OPERATOR key and + ' softkey.
To
display " / ", press
the MATH OPERATOR
key and / ' softkey
To display " ( ", press the MATH OPERATOR key and ( ' sollkey,
To display " ) ", press the MATH OPERATOR key and ) ' softkey.
To
display any alphabet key, let the LED indicator in the blue shiftkey
light, and press any alphabet key.
14. Sequentially press the PROGRAM key and
the
'QUIT editor' and 'RUN' soft!.
Wait until the O P ' softkey changes to. intensified green.
15, Press the SCALE REF key and the 'A AUTO SCALE' softkey for auto scaling.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com