PERFORMANCE TESTS
2, Sequentially press the following HP 4195A front panel keys,
CONFIG, 'NETWORK', PRESET
AMPLITUDE (CHANNEL 1 side), 1 , 0 , ENTER/EXECUTE,
REF ATTEN (CHANNEL 1 side), 1 , 0 , ENTER/EXECUTE
TEST ATTEN (CHANNEL i Side), 1 , 0 , ENTER/EXECUTE
START, 1 , 0 , 0 kHz/dBrn
RES BW, 1 , kHz/dBm
MENU, 'TRIGGER menu', 'SINGLE mode', 'return'
'TYPE lin tog'
DISPLAY, 'TRACE B on oft'
(log" In the 'TYPE lin log' sioltkey will change to
Intensified green.)
("off" In the 'TRACE B on oft' sottkey will change
to intensified green,)
NOTE
If you are testing the HP 41952B, skip step 3.
3. Press tha following keys, and skip step 4,
STOP, 2 , 0 , 0 , MHz/V
4, Press the following keys (only for testing the HP 41952B).
STOP, 5 , 0 , 0 , MHz/V
CAL, 'more 1/2', 'ZO 509 75f2',
( " M ' in the '7.0 Bon 7551' softkey will change to intensified ween.)
5. Connect the N(m) SHORT standard to the HP 41952A/B's TEST PORT connector.
6. Press the TRIG/RESET key on the HP 4195A's front panel, to makeia single sweep
measurement.
7. Enter the HP 4195A "C=A" command using the following keystrokes.
CLR LINE, blue shiftkey, C , = , A , ENTER/EXECUTE
8. Connect t h e N(m) LOAD standard ( H P 909C Opt, 200, Opt. 012 f o r testing
HP 41952A, HP 85036B's 75 CI standard (85036-60008) for testing HP 41952B) to the
HP 41952A/B's TEST PORT connector.
9. Press the HP 4195A's TRIG/RESET key.