9.
PERFORMANCE TEST'S'
2, Toni the HP 4195A.OFF and ON .to Initialize It„-and then sequentially press the fol-
, „ t
•
lowIng)HP 4195A front panelkeys;
CONF1G, 'S-PRMTR', 'S11', PRESET
AMPLITUDE (CHANNEL 1,sIde),1,, 1 ENTER/EXECUTE
REFi'ATTEN (CHANNEL, I side), 1 , 0 , ENTER/EXECUTE
TEST'IMEN (CHANNEL 1 side), 1 0 ,ENTER/EXECUTE
START, 1,, 0 , 0 , ktiz/dBm, ' •
STOP, 51,10 , 0 , MHz/V
RES. BW, 1 , kHz/dBm•
•
MENU, 'TRIGGER menu', 'SINGLE,mode'
DISPLAY, 'TRACE B on off':.'
("off' iri the 'TRACE B on off' softkey will change
to intensified green.)
3. Perform the following procedure (steps a through h), to perform a 1 PORT FULL
CALIBRATION for the TEST' PORT connector of the HP 41952A (test equipment).
a. Sequentially press the HP 4t95A's CAL key, and the `S-PRMTR CAL menu' and
`ONE PORT FULL CAL' softkeys. '
)
b. Connect the HP 85032B's N(m) OPEN standard to the TEST PORT connector of
the HP 41952A (test equipment).
c. Press the 'OPEN' softkey end the ENTER/EXECUTE key. Wait until the single
sweep measurement is completed.
d. Connect the HP 85032B's N(m) SHORT standard to the TEST PORT connector
of the HP 41952A (let equipment).
e. Press the 'SHORT' softkey and ENTER/EXECUTE key. Wa i t until the single
sweep measurement is completed.
f. Connect the HP 85032B's N(m) LOAD standard to the TEST PORT connector of
the HP 41952A (test equipment).
Press the 'LOAD' softkey and the ENTER/EXECUTE key. Wait until the single
sweep measurement is completed.
h. Press the CAL key and the 'CORRECTN on off' softkey. The "on" In this softkey
will change to Intensified green.
4-14