2. Set the network analyzer's start frequency to 10 MHz and its
stop frequency to 600 MHz,,,
3. Set the network analyzer's output level to 0 dBm.
4. The network an•qier's magnitude ratio reading should be
approximately 0 dB.
5. Conned a termination (50 CI termination fiiir the HP 41052A or
75 CI termination for the HP 41952B), to the HP 41952A/B's
TEST PORT connector.
6. The network analyzer's magnitude ratio reading shquld be
less than -30 dB,
3.5: