![HP HP 41952A Operation Note Download Page 35](http://html.mh-extra.com/html/hp/hp-41952a/hp-41952a_operation-note_166195035.webp)
PERFORMANCE TESTS
2, Sequentially press the lollowing HP 4195A front panel keys,
CONFIG, 'S-PRMTR', '511', PRESET
AMPLITUDE (CHANNEL 1 side), 1 , 0 , ENTER/EXECUTE
REP ATTEN (CHANNEL 1 side), i ; ENTER/EXECUTE
TEST ATTEN (CHANNEL I side), 1 , 0 , ENTER/EXECUTE
START, 1 , 0 , 0 , kHr/dBm
STOP, 5 , 0 , 0 , MHz/V
RES BW, 1 , kHz/dBm
MENU, 'TRIGGER menu', 'SINGLE mode'
.'TYPE lin log'
DISPLAY, 'TRACE B on off'
'return'
(log" in the 'TYPE lin log' soltkey will change to
intensified green.)
("off" in the 'TRACE B on off' softkey will change
to intensified green.)
NOTF
If you are testing a HP 41952A, skip step 3.
3. Press the following keys (only when testing a HP 41952B).
CAL, more 1/2 , 'ZO 50Q 7511'
(the "75cr in the 'ZO 50f1 75cr soltkey will change to intensified green)
4. Sequentially press the HP 4195A's CAL key, and the 'S-PRMTR CAL menu' and 'ONE
PORT FULL CAL' soltkeys.
5. Connect the N(m) OPEN standard to the HP 41952A/B's TEST PORT connector.
3. Press the 'OPEN' softkey and the ENTER/EXECUTE key. Wait until the single sweep
measurement Is completed,
7. Connect the N(m) SHORT standard to the HP 41952A/B's TEST PORT connector.
8. Press the 'SHORT' sollkey and the ENTER/EXECUTE key. Walt until the single
sweep measurement is completed.
9. Connect the N(m) LOAD standard to the HP 41952A/B's TEST PORT connector.
110. Press the 'LOAD' soRkey and the ENTER/EXECUTE key. Wait until the single sweep
measurement is completed.
4-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com