![HP HP 41952A Operation Note Download Page 27](http://html.mh-extra.com/html/hp/hp-41952a/hp-41952a_operation-note_166195027.webp)
SECTION 4
Equipment
i
Recommended Model
Qty.
Requirements
,
Network Analyzer
No substitute
'HP 4195A
T/R Test Set
HP 41952A
,
I
1
No substitute
Calibration Kit
i
1
No substitute
HP 85032B
(use open/sho'tI50 (1 termination)
Calibration Kit
i
I
1
Na substitute
HP 85036B ' I
(use open/short/75 it termination)
Termlnatitin
1
Type N(m), 50 C
HP 909C•Opt.200, 0p1012 ' 2
Frequency: S 200 MHz
I
Return Loss: 5 2 dB
Termination
1
Type N(m), 50 fl
HP 909F Opt.012 ' 2
Frequency: 100 kHz to 500 MHz
Return Loss: Z 47 dB
Adapter
1
N(f)-N(f), 50 ft
1
N(m)-N(m), 50 SI
PERFORMANCE TESTS
111=11111111111.1.111111IN
INTRODUCTION T h i s section provides the performance test procedure to ensure
that the HP 41952A/B meets the specifications listed in Table 1-3,
Specifications. The performance test can be performed without
accessing the interior of the HP 41952A/B.
The test results should be recorded into the Performance Test
Record located at the end of this section.
EQUIPMENT REQUIRED
The equipment required for performance testing is listed in Table
4-1. Substitutions can be made for the equipment in this list if
the specifications meet or exceed the specifications listed in the
Requirements column.
Table 4-1, Recommended Test Equipment
Note: U s e d for performance testing the HP 41952B.
2 Used for performance testing the HP 41952A.
4-1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com