Code Message
Cause
- 230
Data corrupt or stale
Data in mainframe memory is fetched after a command
(e.g., MEASure, READ?) has returned data to the
output buffer.
- 231
Data questionable
Resolution is too great for specified range.
Measurement is still taken.
- 240
Hardware error
Hardware error detected during power-on cycle. Return
multimeter to Hewlett-Packard for repair.
- 270
Macro error
Deleting a macro that does not exist.
- 273
Illegal macro label
Macro name begins with a character other than a letter.
- 276
Macro recursion error
The macro called is nested too deep within other macro
calls.
- 277
Macro redefinition not allowed
A macro with that label already exists.
- 300
Device specific error
Severe overload, protection relay is opening (caused
by fixed range and severe overload or COM not
connected to Lo). To fix, use autorange, higher range,
or connect COM to Lo.
- 310
System error
Out of macro memory.
- 350
Too many errors
The error queue is full as more than 30 errors have
occurred.
- 410
Query interrupted
Data is not read from the output buffer before another
command is executed.
- 420
Query unterminated
Command which generates data not able to finish
executing due to a multimeter configuration error.
- 430
Query deadlocked
Command execution cannot continue since the
mainframe’s command input
+ 1000
Out of memory
Not enough memory to store the number of
measurements requested.
+ 2000
Invalid card number
There is no multiplexer which corresponds to the card
number specified in the channel list.
+ 2001
Invalid channel number
The channel or range of channels specified for the
(@channel list) parameter does not exist.
+ 2008
Scan list not initialized
Occurs when INIT is executed after error 2000 or 2001
occurred.
+ 2009
Too many channels in channel list
The number of channels in the specified channel list is
larger than the number of channels on the
multiplexer(s).
+ 2012
Invalid channel range
Specifying a channel range such as (@115:100).
+ 2600
Function not supported on this card
Function specified cannot be measured (specifying a
2-wire resistance measurement without a multiplexer,
for example).
+ 2601
Channel list required
Measurement function specified requires a channel list
(specifying a thermocouple measurement without a
multiplexer, for example).
+ 2602
Timer too fast
Aperture time is longer than sample rate.
198 HP E1326B/E1411B Multimeter Error Messages
Appendix B