The channel list is of the form
(@ccnn)
,
(@ccnn,ccnn)
, or
(@ccnn:ccnn)
, where
cc
= card number and
nn
= channel number (105 is channel 05 of card number 1,
for example).
•
To select autorange, specify
AUTO
(or
DEF
) for range or do not specify a value
for the parameter. In the autorange mode, the multimeter samples the input signal
before each measurement and selects the appropriate range.
•
To specify a
MIN
or
MAX
resolution while autoranging, you must specify
CONF:RES AUTO
or
CONF:RES DEF
(you cannot omit the range parameter).
This prevents the
MIN
or
MAX
resolution from being interpreted as a range setting.
•
The fastest aperture time available when autoranging is 100
µ
s. In order to specify an
aperture time of 10
µ
s, you must select a fixed range.
•
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making 2-Wire Ohms Measurements
CONF:RES 1320,MAX,(@105:109)
!Function: 2-wire ohms;
range selected: 1861
Ω
;
MAX resolution: 125 m
Ω
;
specify channel list.
TRIG:COUN 3
!Scan channel list 3 times
(take 4 readings per trigger).
INIT
!Place multimeter in wait-for-trigger state;
store readings in mainframe memory; trigger
source is IMMediate by default.
FETC?
!Place readings in output buffer.
enter statement
!Enter readings into computer.
:TEMPerature
CONFigure:TEMPerature <
transducer
>,<
type
> ,<
channel_list
>
selects the temperature function. All measurements are returned in degrees celsius.
The following transducers can be measurements using the multimeter:
Thermocouples
Thermistors (2-wire or 4-wire measurement)
RTDs (2-wire or 4-wire measurement)
Two-wire temperature measurements can only be made using the scanning
multimeter (a channel list is required).
Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
<
transducer
>
discrete
TCouple | THERmistor |
FTHermistor | RTD | FRTD
none
129 Command Reference
Chapter 5