Timer/Pacer
Module Size/Device Type:
Timer Range: 76
µ
sec to 65.5 msec
B, register-based (E1326B)
Resolution: 2
µ
s
C, register-based (E1411B)
Accuracy: 0.01%
Connectors Used: P1
Programmable Delay
No. Slots: 1 (standard terminal panel takes 2 slots)
Delay Range: 40
µ
sec to 16 sec
Resolution: 2
µ
sec
Accuracy: 0.01%
VXIbus Interface Capability:
Slave, interrupter, A16, D16
External Trigger
Interrupt Level: 1-7, selectable
Trigger Condition: negative edge
Minimum Pulse Width: 100 nsec
Power Requirements:
Maximum Trigger Rate: 5 kHz
Voltage +5 +12
(fixed range, 10
µ
sec aperture)
Peak module
current, IPM (A): 0.20 0.55
Dyanmic module
current, IDM (A): 0.01 0.01
Typical Reading Storage
Watts/Slot: 8.5 (E1411B) 4.2 (E1326B)
HP 75000
Number of
Mainframe
Readings
Cooling/Slot:
0.14 mm H
2
O @ 0.71 liter/sec (E1411B)
Series B with
0.07 mm H
2
O@ 0.35 liter/sec (E1326B)
standard memory
50,000
Humidity: 65% 0
°
to 40
°
C
Series B with
Operating Temperature: 0
°
to 55
°
C
512 kB memory
(HP E1300/01A Opt. 010)
100,000
Storage Temperature: -40
°
to 75
°
C
Series B with
EMC, RFI, Safety:
1 MB memory
meets FTZ 1046/1984, CSA 556B.
(HP E1300/01A Opt. 011)
200,000
IEC 348, UL 1244
Series C with
Command Module
(HP E1406A)
100,000
Series C plus E1406A
with Opt. 010
200,000
196 HP E1326B/E1411B Multimeter Specifications
Appendix A