Saving Multimeter Configurations
To minimize repeated programming, up to 10 stand-alone multimeter
configurations can be saved in mainframe/command module memory.
The information saved includes:
Measurement System Parameters
–
Measurement function
–
Range
–
Resolution
–
Aperture time
–
Integration time
–
Autozero
–
Offset compensation
Trigger System Parameters
–
Trigger source
–
Trigger count
–
Trigger delay
–
Sample count
–
Sample source
–
Sample timer
Because channel lists are not included, only stand-alone multimeter
configurations are saved. A configuration is identified by a number from
0 to 9. The configuration(s) remains in memory until power is cycled.
How to Save and
Recall a
Configuration
Multimeter configurations are saved and recalled with the commands:
*SAV register
*RCL register
where register is a number from 0 to 9. The following program shows how
a configuration can be saved and recalled.
114 Understanding the HP E1326B/E1411B Multimeter
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com