Category
1
F
ailures
-
Dead
or
No
Response
,
No
Display
,
Unreadable
Display
,
or
Error
Messages
on
the
Display
This
category
of
failure
usually
involves
one
of
the
following:
a
defective
A6
power
supply
a
defective
A2
CPU
P
oor
contact
with
the
rear
plane
connector
board
A
digital
failure
in
any
board
assembly
A
defective
display
Troubleshooting
A6
P
ower
Supply
Assembly
Problems
Begin
here
if
the
problem
appears
to
be
power
supply
related.
Check
the
Rear
P
anel
LEDs
Switch
the
analyzer's
line
power
ON.
Check
the
condition
of
the
two
LEDs
visible
through
the
hole
in
the
rear
panel
to
the
right
of
the
A
C
power
plug.
If
the
green
LED
is
on
and
the
red
LED
is
o:
This
represents
a
normal
condition.
Continue
with
\Measure
the
P
ower
Supply
V
oltages"
to
conrm
that
the
power
supplies
are
functioning
properly
.
If
neither
LED
is
on:
Check
the
line
fuse
,
located
in
a
holder
above
the
A
C
power
jack,
and
replace
it
with
the
spare
provided
if
necessary
.
Check
the
A
C
power
line
.
If
the
LEDs
are
still
o,
replace
the
power
supply
.
If
the
green
LED
is
on,
the
red
LED
is
on
steadily
,
and
the
fan
is
o:
Check
the
line
voltage
selector
switch,
located
below
the
A
C
power
jack.
If
it
is
correct,
remove
the
power
supply/display
assembly
and
make
sure
that
the
cable
from
the
on/standby
switch
to
the
power
supply
assembly
is
connected.
With
the
power
supply/display
assembly
outside
the
analyzer
,
plug
it
in
and
switch
the
analyzer's
line
power
ON..
If
the
fan
is
still
o:
Check
the
four
green
power
supply
LEDs
visible
from
the
right
side
of
the
display
enclosure
(as
viewed
from
the
front).
If
the
LEDs
are
on,
the
fan
is
probably
broken.
Replace
the
fan
or
the
entire
power
supply
assembly
.
If
the
green
LED
is
on,
the
red
LED
is
on
steadily
,
and
the
fan
is
on:
Go
to
\Measure
the
P
ower
Supply
V
oltages
."
Note
that
it
is
normal
for
both
the
green
and
red
LEDs
to
be
on
when
the
power
switch
is
in
ST
ANDBY
,
but
not
when
the
switch
is
in
the
ON
position.
If
the
green
LED
is
on,
and
the
red
LED
is
blinking:
The
power
supply
is
in
shutdown.
Remove
the
power
supply/display
assembly
from
analyzer
,
plug
it
in
and
switch
the
analyzer's
line
power
ON.
T
roubleshooting
and
Block
Diagrams
4-5
Summary of Contents for 87114B
Page 7: ......
Page 19: ...Absolute Power Accuracy broadband Dynamic Accuracy narrowband 1 4 Performance T ests ...
Page 39: ......
Page 73: ......
Page 89: ......
Page 115: ......
Page 116: ......
Page 118: ...Figure 5 1 The Service Key Menus 5 2 Service Related Menus ...
Page 132: ......
Page 162: ......
Page 164: ......
Page 170: ......