![HP 87114B Service Manual Download Page 60](http://html.mh-extra.com/html/hp/87114b/87114b_service-manual_161524060.webp)
Note
If
you
must
cycle
the
analyzer's
line
power
,
and
you
did
not
store
the
correction
constants
to
disk
before
starting
this
adjustment,
you
need
to
repeat
all
of
the
adjustments
done
up
to
this
point,
then
repeat
the
source
power
correction.
If
you
did
save
the
correction
constants
to
disk,
recall
the
correction
constants
data
from
disk
(as
described
in
\Storing
and
Recalling
Correction
Constants"),
then
go
back
to
step
1
to
repeat
this
adjustment.