Figure
2-2.
TP11
and
L162
for
Fractional-N
V
CO
A
djustment
5.
Use
a
nonconductive
adjustment
tool,
such
as
a
small
ceramic
or
plastic
atblade
screwdriver
,
to
adjust
L162
(see
Figure
2-2)
until
TP11
reads
01.17
60.03
volts
with
respect
to
board
ground.
Remove
the
adjustment
tool
and
measure
the
voltage
again
to
insure
that
it
has
not
changed.
6.
If
you
are
going
to
make
the
fractional-N
spur
adjustment,
continue
to
that
procedure
without
changing
the
equipment
setup
.
If
you
are
not
going
to
make
the
fractional-N
spur
adjustment,
reassemble
the
analyzer
.
Adjustments
2-5