D
W
avelength
Repeatability
1.
Set
the
wavelength
meter:
a.
Press
SYSTEM
4
Preset
5 .
b.
Press
DISPLA
Y
4
Peak
WL
5 .
c.
Press
DEL
T
A
4
OFF
5 .
d.
Press
MEASUREMENT
4
Cont
5 .
2.
Set
the
TLS
(DUT)
a.
Switch
o
Coherence
Control.
b.
Set
the
output
power
to
-10dBm
(8168F
to
-7dBm),
activate
laser
output.
c.
Set
the
wavelength
to
HP
8167A
1330nm
HP
8168D
1565nm
HP
8168E
1575nm
HP
8168F
1575nm
3.
Set
the
TLS
to
any
wavelength
in
its
range
and
note
it
in
the
test
record.
4.
Set
the
TLS
back
to
the
wavelength
setup
in
list
item
3
and
wait
a
few
seconds
.
5.
Measure
the
wavelength
with
the
W
avelength
Meter
and
note
the
result
in
test
record.
6.
Repeat
list
item
3
to
list
item
5
with
dierent
wavelength
settings
on
the
TLS
4
times
.
7.
Set
the
wavelength
of
the
8168F
to
1595nm
and
repeat
steps
list
item
3
to
list
item
6
.
D-6
P
erformance
T
ests
Summary of Contents for 8167A
Page 2: ......
Page 4: ......
Page 26: ......
Page 110: ......
Page 130: ......
Page 150: ......
Page 164: ...D Figure D 1 Test Setup for the Wavelength Tests D 4 Performance Tests ...
Page 177: ...D Figure D 5 Test Setup for the Source Spontaneous Emission Test Performance Tests D 17 ...
Page 240: ......