The HP 7090 is shipped from the factory with the line voltage set to the
proper value for the area specified as the shipment's destination. The
voltage selected for the instrument is identified on the rear panel. The line
voltage may be changed by qualified service personnel only_
The HP 7090 is factory equipped with a line fuse appropriate to the factory
set line voltage. Only specified fuses should be used for replacement. To
change or inspect the fuse, turn the fuseholder on the rear panel counter
clockwise until the fuseholder releases. Remove the fuseholder and insert a
"slow-blow" fuse that corresponds with the appropriate voltage setting, as
listed below. Place the fuseholder back into the fuse socket and, while
pressing in, turn the fuseholder clockwise until the lock engages .
Voltage
Fuse
Rating
HP Part
Number
100
V
or 120
V
2.5 AT (SB)/250
V
2110-0015
220
V
od40
V
1.25 AT (SB)/250
V
2110-0305 (U.S.)
or
211()'0472 (Emopean)
WARNING
To avoid the possibility of injury, disconnect the ac power cord before
installing or replacing a fuse .
NOTE:
Fuses and fuse caps appropriate
to
the instrument shipment des
tination are installed at the factory.
U.8.
fuses are
% X Ill.!
inches and use
HP 2110-0565 fuse caps. European fuses are 5
X
20 millimetres and use
HP 2110-0567 fuse caps
.•
The power cord provided with the HP 7090 was chosen to be compatible
with the line power outlets available at the shipment destination. The cord's
female socket, which connects with the power-input plug on the instrument,
is standard on
all
cords. The male plug, which inserts into your line power
outlet, is configured for the shipment destination. If your HP 7090 has the
wrong power cord for your requirements, please contact your local HP
Sales and Support Office. The following chart shows the available power
cord plug options and their HP part numbers.
Fuse Protection
Power Cord
Owner's Information 1-3