Scans by ArtekMedia © 2007
llUlide-WirnWw Triggering
(Buffered Recording Mode)
Buffered llUlide-Window
and
Pre-Trigger Daro.
Because the buffers contents are filled following the required sequence of
inside-window data, post-trigger times are not applicable and, if specified
for a buffered recording, are ignored by the HP
7090_
Pre-trigger times may
be specified, however.
If you select a pre-trigger time, the required number of consecutive inside
window samples is reduced accordingly. as illustrated. below. Pre-trigger
data consists of samples that immediately precede the buffered sequence
of inside-window samples. Note that the buffers are filled when the
Channel ! input signal has remained inside the window for a time equal
to
TOTAL TIME - PRE TRIGGER
time.
1000 SAMPLES
I
��:���L ,
SAMPLES
SAMPLES
PRE.TR'GGER
I
INSIDE WINDOW
I
-
- - - - - - - - - - - - TRIGGER OCCURS
FROZEN
5-8
Trigger Modes and Operations
�
�
�
•
�S
��