Scans by ArtekMedia © 2007
.UUP CO
•
•
•
•
•
•
•
•
•
•
•
The following description of right front-panel controls and indicators is
numerically keyed to the accompanying figures.
1.
Setup Condition Selectors and Indicators:
Used to select setup
condition choices. The current value of any selected condition is dis·
played on the LeD and may be altered using the control wheel. The
following setup conditions are available:
•
range
- sets channel input range and sensitivity .
•
offset
- offsets zero position of input signal.
•
grid divisions
- specifies the number of grid divisions to be drawn
on the recording area's
X-
and V-axes.
•
total
t
im
e
- specifies recording time.
•
post-trigger/pre-trigger
- sets delay time between trigger and
recording; sets amount of pre-trigger buffered recording.
•
trigger level
- specifies Channel
1
voltage level required to trigger
recordings; specifies center voltage of trigger windows.
•
trigger width
- specifies voltage width of trigger windows.
•
time and date
- sets internal real-time clock (accessible with
SHIFT
push button only).
2.
Control Wheel:
Rotated to establish setup condition values.
3.
PAPER SIZE
Push button:
Used to select paper size. Toggles between
8.5 X 11
and
1
1
X 17
inches (with rear· panel paper format switch set to
ANSI)
or between
A4
and
A3
(with rear-panel paper· format switch set
to ISO. Current paper size selection is displayed on the LCD.
(SHIFT) LOCAL:
Used to restore the
HP 7090
to local mode following a
computer·issued RLl (remote) instruction, which "locks out" the front·
panel controls during remote applications.
4.
RESTORE SETUP
Pushbutton:
Used to restore from nonvolatile memory
the set of front-panel setup values and modes stored previously using
the
STORE
push button.
(SHIFT) SAVE SETUP:
Stores the currently selected front-panel setup values
and modes in nonvolatile memory.
5.
SHIFT
Pushbutton:
Accesses secondary functions (Iabeled in blue) of
dual-function pushbuttons. Toggles between active
("SHIFT"
displayed
on LCD before dual·function pushbutton is pressed) and inactive
("SHIFT"
not displayed before dual-function push button is pressed).
6.
COARSE
Push button:
Permits coarse, rapid tuning of setup condition
values. Toggles between active ("COARSE" displayed on LCD) and
inactive ("COARSE" display extinguished, permitting fine-tuning with
control wheel).
7.
Trigger Mode Selector and Indicators:
Used to select trigger mode.
2-6
Instrum
en
t Description and General Operation