Plum XL
CHECK
SETTINGS
Rate or VTBI not set
Turn to SET RATE or
SET VTBI to check
setting or enter values
DOOR/
CASSETTE
Cassette door open
Cassette
improperly loaded
Cassette
improperly primed
Cassette failed
valve leak test
Turn to OFF CHARGE,
close cassette door,
then restart
Turn to OFF CHARGE,
reload cassette, then
restart
Turn to OFF CHARGE,
reprime cassette, then
restart
Turn to OFF CHARGE,
open and close cassette
door, then restart. If
condition recurs,
replace PlumSet
LOW BATTERY
Approximately 30
minutes of battery
power remains
Connect to AC power
Note:
Pressing the
[SILENCE] key mutes
the audible alarm for 15
minutes from the time
the LOW BATTERY
alarm occurred
Note:
When the battery
discharges, pumping
stops and the alarm
sounds continuously for
one minute before the
device shuts down
completely
OCCLUSION
Clamp closed
Tubing kinked
Possible clotted
catheter
Open clamps
Unkink tubing
Check IV site
MESSAGE
POSSIBLE CAUSE
CORRECTIVE ACTION