System Operating Manual
27
7.2
Malfunctions
During a malfunction, an "Er" and an error number display on
the screen and the audible alarm sounds. To verify the
malfunction, proceed as follows:
Record the error number displayed
on the screen.
Turn the control dial to OFF
CHARGE.
Turn the control dial to its previously
set setting to retest the device. If the
malfunction recurs, remove the
device from service.
TURN TO RUN
Control dial not in
OFF CHARGE or
RUN position and
no key is pressed
for five minutes
Turn control dial to RUN,
OFF CHARGE, or
HOLD/RESET
VTBI
COMPLETE
Secondary Only, or
Primary Only:
Programmed VTBI
completed
Piggyback:
Primary
VTBI completed
Note:
KVO also
displays on the
screen indicating
the 1mL/hr KVO
rate is in progress
Discontinue infusion, or
change container and
program new VTBI
setting
MESSAGE
POSSIBLE CAUSE
CORRECTIVE ACTION