System Operating Manual
23
Primary only:
Remove the container
attached to the secondary inlet port,
then cap the port (if appropriate).
To resume delivery, turn the control
dial to RUN.
6.5
Changing Containers
To change a container, use aseptic technique and proceed as
follows:
Turn the control dial to
HOLD/RESET.
With the cassette door closed, spike
the new container.
Note:
If opening the cassette door,
close the primary and secondary
clamps before removing containers
(to prevent mixing).
Turn the control dial to SET VTBI,
then set the volume to be delivered.
OFF
CHARGE
SET RATE
SET VTBI
RUN
HOLD/RESET
CLEAR
VOL
OFF
CHARGE
SET RATE
SET VTBI
RUN
HOLD/RESET
CLEAR
VOL
OFF
CHARGE
SET RATE
SET VTBI
RUN
HOLD/RESET
CLEAR
VOL