13.0 SPECIFICATIONS
PHYSICAL:
Dimensions:
Approximately 8.25H x 7.5W x 8.75D
inches (including pole clamp)
Weight:
Approximately 7.75 lbs (with battery)
Casing:
High-impact plastic
ELECTRICAL:
Power Requirements:
100-130 VAC, 47/63 Hz, less than
35W
Power Cord:
Hospital-grade AC cord, 10 ft long,
with transparent plug
Fuses:
0.5 A, 250 V, Slow Blowing
Battery:
Sealed lead-acid, rechargeable 8 V
battery, internal to device. Accessible
for ease of field replacement, with
color-coded leads and polarized
connector
Battery Life:
A fully charged new battery will
provide approximately eight hours of
operation at 125 mL/hr or 1000 mL
total volume delivered, whichever
occurs first. The device should be
operated on battery power for six
continuous hours at least once every
six months for optimum performance
and battery life