System Operating Manual
9
4.2
Keys
[PRI-SEC] key selects the fluid line to
program. Press the [PRI-SEC] key
when the control dial is in SET RATE
or SET VTBI to toggle between the
primary and secondary line.
[TITRATE] key adjusts the fluid
delivery rate up or down while
pumping is in progress. Hold the
[TITRATE] key while pressing the
or
key to increase or decrease the
delivery rate.
[BACKPRIME] key clears any air
accumulated in the cassette. Press
the [BACKPRIME] key when the
control dial is in HOLD/RESET to
pump fluid from the primary line and
expel the air into the secondary line.
Backpriming is also used to reprime
empty secondary tubing above the
cassette.
[SILENCE] key temporarily mutes
audible alarms. The alarm display and
the LCD screen continue to flash. The
audible alarm resumes after two
minutes if the alarm condition is not
corrected. To silence a Low Battery
alarm, refer to the alarm tips in the
Troubleshooting Section.
4.3
AUDIO SWITCH
The audio switch has two sound level
settings. Turn the switch toward the
front panel for the LOW setting or
toward the back for the HIGH setting.
PRI
SEC
TITRATE
BACK
PRIME
SILENCE