6. Troubleshooting
90
Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
USING EXT BATT
Y
V
This is a normal
response when:
•
The pump has
been powered on
with external
batteries as a
power source;
or
•
The pump has
switched to
external battery
power due to a
loss of AC mains
power;
or
•
The pump has
switched to
external battery
power while
running on
internal batteries.
When powering on the pump:
•
Press [
YES/ENTER
] to
confirm external battery
power source.
When switching from AC Mains
power:
•
Press [
SILENCE
] to clear the
alarm.
CAUTION: Ensure two
fresh, disposable AA
batteries are installed for
backup power when
using other power
sources (e.g., AC Mains
adaptor, docking station,
battery pack). The pump
will continue to operate
on backup power if the
external power source
fails. Refer to the caution
regarding Power Loss on
page 13
Alarm Display
Message
Is the
pump
infusing?
Alarm Cause/
Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
430-10984-001.book Page 90 Friday, March 31, 2006 4:49 PM