6. Troubleshooting
94
Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Other Display Messages
Other Display
Message
Cause/Condition
Remedy
KEYPAD LOCKED
User has attempted to
perform a function that has
been locked out.
Unlock the keypad to access the
desired function.
NOT ALLOWED
DURING INFUSION
PRESS STOP TO
HALT DELIVERY
User has attempted to
perform a function that is
not available in
RUN
mode.
•
Press [
STOP
] to access the
desired function.
•
Press [
YES/ENTER
] to return to
the
RUN
mode display.
ROUNDING
•
The pump sounds
four quick beeps
and displays the
rounded number.
The pump has rounded a
number that was entered
during programming.
Press [
YES/ENTER
] to accept the
rounded value and continue to the
next step.
DOSE IN PROGRESS
NEW CONTAINER
NOT ALLOWED
User has attempted to
access NEW CONTAINER
function when it is not
allowed.
1
Press [
STOP
].
2
Increase the container size to
complete the dose.
3
Set
NEW CONTAINER
through the Change Menu.
CANNOT CHANGE
CLOCK WHILE TIME
BASED DELIVERY
IS IN PROGRESS
User has attempted to
change the clock during a
time-based delivery.
1
Press [
STOP
].
2
Clear the program through the
Change Menu.
3
Change the clock through the
Options Menu.
4
Reprogram the pump.
THIS OPTION NOT
AVAILABLE WHILE
PROGRAMMING THE
PUMP
User has attempted to
access a function that is not
available from the first
programming display.
1
Press [
YES/ENTER
] to return to
the first programming display.
2
Complete programming.
3
Access desired function.
430-10984-001.book Page 94 Friday, March 31, 2006 4:49 PM