Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
93
6. Troubleshooting
NO INTERNAL BATT
NO POWER LOSS
ALARM AVAILABLE
ENTER TO CONFIRM
N
A/V
No disposable batteries
or dead disposable
batteries; AND external
power.
1
Press [
YES/ENTER
] to
confirm.
2
Power-off the pump and
install two fresh,
disposable AA batteries
as a backup power
source.
CAUTION:Ensure
two fresh, disposable
AA batteries are
installed for backup
power when using
other power sources
(e.g., AC Mains
adaptor, docking
station, battery pack).
The pump will
continue to operate
on backup power if
the external power
source fails. Refer to
the caution regarding
Power Loss on
page 13.
PROGRAMMING
INCOMPLETE
N
A/V
Programming has not
been completed and no
keys have been pressed
within three minutes.
1
Press any key to clear
the alarm.
2
Complete the program
entries.
START
•
START flashes on
the display. Other
parts of the display
are not affected.
N
A/V
A Stop Mode Callback
has occurred because the
programmed pump has
remained in the STOP
mode for more than three
minutes.
1
Press [
SILENCE
] to
quiet the alarm for 3
minutes.
2
Follow the instructions
on the display to clear
the alarm.
Alert Display Message Is the
pump
infusing?
Alar
m
Cause/Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
430-10984-001.book Page 93 Friday, March 31, 2006 4:49 PM