5. Clinician Instructions
78
Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Clinician Locking Sequence Instructions
Locking the Pump
To lock the pump from the
RUN
or
STOP
mode:
1
Press [
OPTIONS
] to display the Options Menu.
2
Select
KEYPAD LOCK
from the Options Menu.
3
Select the desired lock level. For
CONTAINER
and
FULL
Lock, go to step 5.
4
For
RATE CHANGE
Lock, set the minimum and maximum programmable ranges
for the continuous delivery rate and bolus dose in a Pain Management program.
•
MIN
flashes on the display. Enter the minimum value.
•
MAX
flashes on the display. Enter the maximum value.
5
Press [
YES/ENTER
] when done.
6
Repeat steps 4 and 5 for a bolus dose if available in the current program.
7
Enter the lock sequence number [
1
][
3
][
0
][
0
][
0
]. Asterisks display as the numbers
are entered.
Unlocking the Pump
1
Press [
OPTIONS
] to display the Options Menu.
2
Select
KEYPAD LOCK
from the Options Menu.
3
Select the lock level to be unlocked.
4
Enter the lock sequence number for that level.
5
The pump displays the current lock level. Press [
YES/ENTER
] to return to the
mode the pump was in when [
OPTIONS
] was pressed.
I
Tip: Press
to toggle between the MIN and MAX values.
G
Note: Default lock sequence numbers are shown.
430-10984-001.book Page 78 Friday, March 31, 2006 4:49 PM