2. Basic Operation
20
Hospira GemStar Manual 430-10984-001 (Rev. 03/06)
Connecting the GemStar® AC Mains Adaptor
1
Connect the cord of the AC mains adaptor to the connector port labeled “3 VDC”
on the bottom of the pump.
2
Plug the AC mains adaptor into a standard wall outlet.
CAUTION: ALWAYS connect to a grounded AC mains power source when
using the AC mains adaptor. USE ONLY AC mains adaptors
specifically labeled for use with the GemStar® Pump.
When removing the cord from the bottom of the pump, firmly
grasp the connector and pull straight out. DO NOT twist or bend
the cord or connector.
Ensure two fresh, disposable AA batteries are installed for backup
power when using other power sources (e.g., AC Mains adaptor,
docking station, battery pack). The pump will continue to operate
on backup power if the external power source fails. Refer to the
caution regarding Power Loss on page 13.
G
Note: The green LED
on the pump remains lit when connected to AC mains
power.