56
6.
Pull the old Keypad Ribbon upwards and out through the hole in the chassis.
7.
Insert and run the Keypad Ribbon through the chassis and route it behind the Control Circuit board.
8.
Tighten or replace all screws to the Control Circuit Board making sure to attach the KeyPad’s
Grounding Wire to the bottom front screw.
9.
Replace the cover.