7.9. Replacing the MAC Valve (50-2237)
PURPOSE:
This work instruction will replace the MAC Valve in a SPE-DEX 4790 Module.
TOOLS:
-
¼ inch Nut Driver
-
5/16 inch Wrench
-
Molex Pin Extraction Tool (99-0766)
-
Small Standard Screwdriver
-
Small Phillips Screwdriver
-
Small Wire Cutters
PARTS REQUIRED:
50-2237 – MAC Valve
99-0766 – Molex Pin Extraction Tool
CAUTIONS:
Prior to beginning, the Controller should be turned off and the Module should have the data/power cable
unplugged. The nitrogen supply should be turned off and the lines vented to atmosphere.
PROCEDURE:
9.
Remove the five (5) screws from the rear of the Module using the ¼ inch Nut Driver.
10.
Remove the cover by holding the bottom of each side and pushing the bottom backwards to
disengage the front locking tabs. Lift the cover off.
11.
Locate the MAC valve and follow the black wires to the Molex connector.
12.
The pin layout for the Molex connector half attached to the valves is shown below. Note that the
locking mechanism is shown up.