Conductivity measurement
40
5. Press the
keys to enter the number of the cell constant of the
conductivity cell and then press the
key to confirm.
To return to the measurement mode, press the
key.
Note
・
The unit used for the cell constant corresponds the unit set in “ Setting the conductivity unit
・
When the cell constant is changed through the cell constant setting, all the previous
calibration data is deleted.
・
Match the unit indicated on the conductivity cell to the unit set in the instrument.
10 m
-1
⇔
0.1 cm
-1
100 m
-1
⇔
1 cm
-1
1000 m
-1
⇔
10 cm
-1
key
Enter a
number