Other settings
71
■
Other settings
●
Printing measured values and calibration data
Print out the measured value or the calibrated value displayed on the instrument, or the
measurement data or the calibration data saved in the instrument. If the repeatability is
inspected, the inspection data is printed out with the calibration data.
Pressing the
key starts printing during displaying the data you want to print. Use the
printer cable to connect the printer unit with the instrument beforehand. If the automatic
data memory is set to "ON", you cannot print out in the measurement mode.
Reference
For details of how to display measurement data and calibration data, refer to the respective
section.
“ Displaying saved data ” (page 52)
“ Displaying the latest calibration and inspection data ” (page 56)
Refer to the following table for the pH (two point calibration) and conductivity.
< Measurement data (CH1)>
Printout format
Description
Date
:
2016/01/01
Measurement date
Time
:
09
:
00
Measurement time
pH
:
5.22
Measured value
(Outside the display range: "UR" or "OR")
HOLD
:
HOLD
HOLD status
Hold value: HOLD
Not hold value: INST
Temperature
:
25.5°C ATC
Temperature value (outside the display
range: "Ur" or "Or") and temperature setting
Inst. model
:
LAQUA-PC1100
Instrument model
Inst. SN
:
KL1TSN20
Instrument serial number
Elect. status
:
OK
Electrode status based on calibration result
Offset
:
0.7 mV
Asymmetry potential of calibration data
Sensitivity
Sensitivity of calibration data
pH 4.01 - 6.86
:
98.9%
Summary of Contents for LAQUA-PC1100
Page 1: ...Benchtop pH ORP Conductivity Meter LAQUA PC1100 Instruction Manual CODE GZ0000427795 ...
Page 2: ......
Page 16: ......
Page 28: ...M E M O 12 ...
Page 94: ...M E M O 78 ...
Page 100: ...M E M O 84 ...
Page 108: ...M E M O 92 ...
Page 117: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com ...