Measurement settings
56
■
Measurement settings
●
Displaying the latest calibration and inspection data
• In the case of displaying the latest pH calibration data and
repeatability inspection data
1. Press the
key to enter the setting mode when "ACT" icon is displayed on the
upper part of the screen.
2. Press the
keys to select "pH" (pH setting) and then press the
key.
3. Press the
keys to select "CAL.DAT" (display calibration data) and then
press the
key.
The electrode status based on the calibration result is displayed. You can change the
display item among "OFS" (asymmetry potential), "SLP" (sensitivity), and "REP"
(repeatability).
When pressing the key, the calibration data is printed.
To return to the measurement mode, press the
key and then press the
key.
key
key
key
key
key
key
Asymmetry potential display
Sensitivity display
Repeatability display
key