Conductivity measurement
47
• Manual calibration
1. Clean the conductivity cell with pure water (or
deionized water) and wipe it with filter paper or
tissue paper.
Do not touch the black electrode part.
Refer to the instruction manual of the conductivity
cell for how to clean the conductivity cell.
2. Immerse the conductivity cell in the standard
solution.
Make sure the hole at the upper part of the cell is
immersed.
3. Press the
key to enter the calibration mode
when "ACT" icon is displayed on the lower
part of the screen.
Stability judgment starts and the COND icon
blinks. When the measured value stabilizes, the
COND icon changes from the blinking state to the
lit state.
4. Press the
keys to enter the electrical conductivity value of the
standard solution used for calibration at the measurement temperature.
5. Press the
key or
key.
Calibration to the set standard solution value at the measured temperature is
performed. The 1st point calibration ends and the calibration history icon
corresponding to the calibration range lights, indicating that 1st point calibration is
complete.
The display returns to the measurement screen.
For multiple point calibration, repeat the steps 1. to 5.
The calibration points are up to 5 points.
Stabilized
key
Enter a
number
key or
key
key
Summary of Contents for LAQUA-PC1100
Page 1: ...Benchtop pH ORP Conductivity Meter LAQUA PC1100 Instruction Manual CODE GZ0000427795 ...
Page 2: ......
Page 16: ......
Page 28: ...M E M O 12 ...
Page 94: ...M E M O 78 ...
Page 100: ...M E M O 84 ...
Page 108: ...M E M O 92 ...
Page 117: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com ...