When an error message appears
87
●
ERROR No.0006
Maximum calibration points exceeded
Attempted to perform 6th point calibration during pH calibration.
●
ERROR No.0007
Cannot identify standard solution
Unable to automatically detect the standard solution during pH calibration.
●
ERROR No.0008
Calibration interval error
・
More than the set number of days has elapsed since calibration was last performed.
・
The calibration interval alarm is "ON" and calibration has not been performed.
Cause
How to solve problem
6th point calibration is
attempted.
Up to five points can be calibrated.
Cause
How to solve problem
There is a problem with the
standard solution.
Use new standard solution.
Settings about the standard
solution do not match the
instrument.
Check if the instrument settings and the specifications of
the standard solution are compatible.
The electrode is dirty.
Clean the electrode.
The electrode is cracked.
Replace the electrode.
The standard solution
concentration is fluctuating.
Replace the internal solution in the reference electrode.
Electrode is not immersed
enough to cover liquid junction.
The electrode must be immersed up to the liquid junction.
As a guide, immerse to at least 3 cm from the tip of the
electrode.
Cause
How to solve problem
Calibration has not been
performed for the set number of
days of the calibration interval or
longer.
Perform calibration.
The calibration interval alarm is
"ON" and calibration has not
been performed.
Perform calibration.
Summary of Contents for LAQUA-PC1100
Page 1: ...Benchtop pH ORP Conductivity Meter LAQUA PC1100 Instruction Manual CODE GZ0000427795 ...
Page 2: ......
Page 16: ......
Page 28: ...M E M O 12 ...
Page 94: ...M E M O 78 ...
Page 100: ...M E M O 84 ...
Page 108: ...M E M O 92 ...
Page 117: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com ...