Deactivating Policy Enforcement
If the AreaRAE Plus/Pro screen displays the message that it must be bump tested or calibrated,
and if the option to bypass bump testing or calibration is not available, you should shut off the
instrument and follow the procedure outlined here if you want to change the Policy
Enforcement settings:
1.
Use a USB cable to connect the AreaRAE to a computer running Safety Suite Device
Configurator (SSDC).
2.
Enter Diagnostic Mode on the AreaRAE Plus/Pro (with the instrument turned off, press
and hold [Y/+] and [MODE] until it starts up.
3.
After startup, enter the password when prompted (default is “0000”) and press [MODE].
4.
Press [N/-] repeatedly until you see the “Enter Communications Mode?” screen.
5.
Press [Y/+] to enter Communications Mode.
6.
Safety Suite Device Configurator (SSDC).
7.
Input the username and password (the default username is “administrator” and the
default password is "Default123").
8.
Click “LOG
IN”
9.
Click on the instrument’s serial number.
10.
Click on the “Settings” tab.
11.
Scroll down to “Policy Settings.”
12.
Deselect Policy Enforcement features you do not wish to use.
13.
Click “Save”
14.
When the upload is done, exit Safety Suite Device Configurator (SSDC).
15.
Press [Y/+] on the AreaRAE Plus/Pro to exit Communication Mode.
AreaRAE
87
User Manual
Summary of Contents for AreaRAE Plus
Page 1: ...AreaRAE WIRELESS MULTI GAS MULTI THREAT DETECTOR ...
Page 2: ......
Page 17: ...AreaRAE 16 User Manual ...
Page 22: ...Temperature Relative humidity AreaRAE 21 User Manual ...
Page 34: ...The instrument should wake up and begin monitoring immediately AreaRAE 33 User Manual ...
Page 145: ...Rev C September 2021 User Manual P N W01 4001 000 ...