Enter Programming in Advanced Mode
1.
To enter Programming Mode, press and hold [MODE] and [N/-] until you see the
Calibration screen. No password is necessary in Advanced Mode.
2.
Press [N/-] to step through the programming screens.
Note: Holding down [N/-] advances quickly through the menu options.
When a menu item is selected, its name is shown at the top of the screen, and its icon is
enlarged.
To enter a menu and view or edit parameters in its submenus, press [Y/+].
AreaRAE
52
User Manual
Summary of Contents for AreaRAE Plus
Page 1: ...AreaRAE WIRELESS MULTI GAS MULTI THREAT DETECTOR ...
Page 2: ......
Page 17: ...AreaRAE 16 User Manual ...
Page 22: ...Temperature Relative humidity AreaRAE 21 User Manual ...
Page 34: ...The instrument should wake up and begin monitoring immediately AreaRAE 33 User Manual ...
Page 145: ...Rev C September 2021 User Manual P N W01 4001 000 ...