l
To calculate the best High value, first add the Blocked and Unblocked values and divide
by 2: (Blocked value + Unblocked value) / 2 = correct Stall High value.
l
Press [N/-] to advance to the next Pump screen.
l
Use the [Y/+] and [N/-] keys to increase or decrease the high value input the results of
your calculation.
l
Press [MODE] to register that value.