Guidance and manufacturer’s declaration –electromagnetic immunity
The WiTouch
®
and WiTouch
®
Pro are intended for use in the electromagnetic environment specified below. The customer or the user of the WiTouch
or WiTouch
Pro should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile RF communications equipment should be used
no closer to any part of the WiTouch or WiTouch
Pro, including
cables, than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
P
d
17
.
1
=
150 kHz to 80 MHz
P
d
17
.
1
=
80 MHz to 800 MHz
P
d
33
.
2
=
800 MHz to 2.5 GHz
where
P
is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and
d
is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as deter-mined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the WiTouch or
WiTouch
Pro is used exceeds the applicable RF compliance level above, the WiTouch or WiTouch
Pro should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the WiTouch or
WiTouch
Pro.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances
between portable and mobile RF communications equipment and the WiTouch and WiTouch
Pro
The WiTouch and WiTouch
Pro are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the WiTouch or WiTouch
Pro can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the WiTouch or WiTouch
Pro as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
P
d
17
.
1
=
P
d
17
.
1
=
P
d
33
.
2
=
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Guidance and manufacturer’s declaration –electromagnetic immunity
The WiTouch
®
and WiTouch
®
Pro are intended for use in the electromagnetic environment specified below. The customer or the user of the WiTouch
or WiTouch
Pro should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile RF communications equipment should be used
no closer to any part of the WiTouch or WiTouch
Pro, including
cables, than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
P
d
17
.
1
=
150 kHz to 80 MHz
P
d
17
.
1
=
80 MHz to 800 MHz
P
d
33
.
2
=
800 MHz to 2.5 GHz
where
P
is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and
d
is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as deter-mined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the WiTouch or
WiTouch
Pro is used exceeds the applicable RF compliance level above, the WiTouch or WiTouch
Pro should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the WiTouch or
WiTouch
Pro.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances
between portable and mobile RF communications equipment and the WiTouch and WiTouch
Pro
The WiTouch and WiTouch
Pro are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the WiTouch or WiTouch
Pro can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the WiTouch or WiTouch
Pro as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
P
d
17
.
1
=
P
d
17
.
1
=
P
d
33
.
2
=
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.