background image

Guidance and manufacturer’s declaration –electromagnetic emissions 

The WiTouch

®

 and WiTouch

® 

Pro are intended for use in the electromagnetic environment specified below. The customer or the user of the WiTouch 

or WiTouch

 

Pro should assure that it is used in such an environment.  

Emissions test  

Compliance  

Electromagnetic environment – guidance  

RF emissions  

CISPR 11  

Group 1  

The WiTouch and WiTouch

 

Pro use RF energy only for its internal 

function. Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.  

RF emissions  

CISPR 11  

Class B  

 The WiTouch and WiTouch

 

Pro are suitable for use in all establishments 

other than domestic and those directly connected to the public low-

voltage power supply network that supplies buildings used for domestic 

purposes. 
  
  

Harmonic emissions 

IEC 61000-3-2  

Not applicable 

Voltage fluctuations /  

flicker emissions  

IEC 61000-3-3  

Not applicable 

 

 

Guidance and manufacturer’s declaration –electromagnetic immunity

 

 

The WiTouch and WiTouch

 

Pro are intended for use in the electromagnetic environment specified below. The customer or the user of the WiTouch or 

WiTouch

 

Pro should assure that it is used in such an environment. 

 

Immunity test  

IEC 60601 test level  

Compliance level  

Electromagnetic environment 

– 

guidance 

Electrostatic discharge 

(ESD) 

 

 IEC 61000-4-2  

±6 kV contact  

 

±8 kV air  

±6 kV contact  

 

±8 kV air

 

 

Floors should be wood, concrete or 

ceramic tile. If floors are covered with 

synthetic material, the relative 

humidity should be at least 30 %.  

 

 

Electrical fast 

transient/burst  

 

IEC 61000-4-4 

 

±2 kV for power supply lines 

 

±1 kV for input/output lines 

±2 kV for power supply lines 

 

±1 kV for input/output lines 

Mains power quality should be that of 

a typical commercial or hospital 

environment. 

 

Surge  

 

IEC 61000-4-5 

 

±1 kV differential mode 

 

±2 kV common mode 

±1 kV differential mode 

 

±2 kV common mode 

Mains power quality should be that of 

a typical commercial or hospital 

environment. 

 

Voltage dips, short 

interruptions and voltage 

variations on power supply 

input lines IEC 61000-4-11  

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 0.5 cycle  

 

40 % 

U

T  

(60 % dip in 

U

T

)  

for 5 cycles  

 

70 % 

U

T  

(30 % dip in 

U

T

)  

for 25 cycles  

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 5 sec  

 

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 0.5 cycle  

 

40 % 

U

T  

(60 % dip in 

U

T

)  

for 5 cycles  

 

70 % 

U

T  

(30 % dip in 

U

T

)  

for 25 cycles  

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 5 sec  

 

 

Mains power quality should be that of 

a typical commercial or hospital 

environment. If the user of the 

[

EQUIPMENT 

or 

SYSTEM

] requires 

continued operation during power 

mains interruptions, it is 

recommended that the [

EQUIPMENT 

or 

SYSTEM

] be powered from an 

uninterruptible power supply or a 

battery.  

Power frequency (50/60 

Hz)  magnetic field IEC 

61000-4-8 

3 A / m 

3 A / m 

 

Power frequency magnetic fields 

should be at levels characteristic of a 

typical location in a typical commercial 

or hospital environment. 

 

NOTE 

U

is the a.c. mains voltage prior to application of the test level. 

 

 

 

 

 

 

Guidance and manufacturer’s declaration –electromagnetic emissions 

The WiTouch

®

 and WiTouch

® 

Pro are intended for use in the electromagnetic environment specified below. The customer or the user of the WiTouch 

or WiTouch

 

Pro should assure that it is used in such an environment.  

Emissions test  

Compliance  

Electromagnetic environment – guidance  

RF emissions  

CISPR 11  

Group 1  

The WiTouch and WiTouch

 

Pro use RF energy only for its internal 

function. Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment.  

RF emissions  

CISPR 11  

Class B  

 The WiTouch and WiTouch

 

Pro are suitable for use in all establishments 

other than domestic and those directly connected to the public low-

voltage power supply network that supplies buildings used for domestic 

purposes. 
  
  

Harmonic emissions 

IEC 61000-3-2  

Not applicable 

Voltage fluctuations /  

flicker emissions  

IEC 61000-3-3  

Not applicable 

 

 

Guidance and manufacturer’s declaration –electromagnetic immunity

 

 

The WiTouch and WiTouch

 

Pro are intended for use in the electromagnetic environment specified below. The customer or the user of the WiTouch or 

WiTouch

 

Pro should assure that it is used in such an environment. 

 

Immunity test  

IEC 60601 test level  

Compliance level  

Electromagnetic environment 

– 

guidance 

Electrostatic discharge 

(ESD) 

 

 IEC 61000-4-2  

±6 kV contact  

 

±8 kV air  

±6 kV contact  

 

±8 kV air

 

 

Floors should be wood, concrete or 

ceramic tile. If floors are covered with 

synthetic material, the relative 

humidity should be at least 30 %.  

 

 

Electrical fast 

transient/burst  

 

IEC 61000-4-4 

 

±2 kV for power supply lines 

 

±1 kV for input/output lines 

±2 kV for power supply lines 

 

±1 kV for input/output lines 

Mains power quality should be that of 

a typical commercial or hospital 

environment. 

 

Surge  

 

IEC 61000-4-5 

 

±1 kV differential mode 

 

±2 kV common mode 

±1 kV differential mode 

 

±2 kV common mode 

Mains power quality should be that of 

a typical commercial or hospital 

environment. 

 

Voltage dips, short 

interruptions and voltage 

variations on power supply 

input lines IEC 61000-4-11  

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 0.5 cycle  

 

40 % 

U

T  

(60 % dip in 

U

T

)  

for 5 cycles  

 

70 % 

U

T  

(30 % dip in 

U

T

)  

for 25 cycles  

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 5 sec  

 

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 0.5 cycle  

 

40 % 

U

T  

(60 % dip in 

U

T

)  

for 5 cycles  

 

70 % 

U

T  

(30 % dip in 

U

T

)  

for 25 cycles  

 

<5 % 

U

T  

(>95 % dip in 

U

T

)  

for 5 sec  

 

 

Mains power quality should be that of 

a typical commercial or hospital 

environment. If the user of the 

[

EQUIPMENT 

or 

SYSTEM

] requires 

continued operation during power 

mains interruptions, it is 

recommended that the [

EQUIPMENT 

or 

SYSTEM

] be powered from an 

uninterruptible power supply or a 

battery.  

Power frequency (50/60 

Hz)  magnetic field IEC 

61000-4-8 

3 A / m 

3 A / m 

 

Power frequency magnetic fields 

should be at levels characteristic of a 

typical location in a typical commercial 

or hospital environment. 

 

NOTE 

U

is the a.c. mains voltage prior to application of the test level. 

 

 

 

 

 

 

Summary of Contents for WiTouch

Page 1: ... on the remote control will flash several times indicating syncing is complete 5 Replace the back cover on the device Note You may need to re sync your remote control and WiTouch device when you replace the battery in the remote control This will ensure communication between your remote control and WiTouch device If your remote control and WiTouch device ever lose communication please repeat these...

Page 2: ...ere or lasts more than five days may indicate a severe condition Stop using the device and consult with your physician Use of the device on the following skin conditions may cause a condition to become worse Do not use the device over or in proximity to these skin conditions abnormal skin skin that is not intact uncleaned unhealthy open wounds rashes swollen red infected inflamed areas skin erupti...

Page 3: ...the Gel Pads to the WiTouch Device The gel pads are intended for single person use They will last depending on skin type oils and pH levels approximately two to five applications Replace the gel pads when they no longer adhere completely Follow these steps to apply the gel pads 1 Separate two gel pads 2 Remove the blue liner from the side being applied to the electrode area Do not remove the green...

Page 4: ...th a comfortable high sensory sensation Treatment Recommendations You can leave the device in position for multiple treatments during the day It will automatically turn Off after two 2 hours of inactivity It is recommended you wait a minimum of 30 minutes between treatments Conducting a Treatment Always read the safety warnings before conducting a treatment Follow these steps to conduct a treatmen...

Page 5: ...covered Battery Replacement When battery replacement is needed the LED will flash as follows To replace the WiTouch Device Batteries 1 Using the included 0 Phillips head screwdriver remove the cover 2 Remove the old batteries and place the new batteries in the correct polarity 3 Replace the battery compartment cover To replace the remote control battery 1 Place the provided screwdriver coin or fla...

Page 6: ...S Hollywog LLC Company warrants the WiTouch Product is free of defects in materials and workmanship This warranty shall remain in effect for 1 year 12 months from the date of purchase If these Products fail to function during the 1 year warranty period due to a defect in material or workmanship at the Company s Option Company or the selling retailer will replace the respective Product without char...

Page 7: ...cial or hospital environment NOTE UT is the a c mains voltage prior to application of the test level Guidance and manufacturer s declaration electromagnetic emissions The WiTouch and WiTouch Pro are intended for use in the electromagnetic environment specified below The customer or the user of the WiTouch or WiTouch Pro should assure that it is used in such an environment Emissions test Compliance...

Page 8: ...ctromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic immunity The WiTouch and WiTouch Pro are intended for use in the electromagnetic environment specified below The customer or the user of the WiTouch or WiTouch Pro should assure that it is used in such an environment Immunity test IEC 60601 te...

Page 9: ...moto destellará varias veces indicando que la sincronización se completó satisfactoriamente 5 Vuelva a colocar la cubierta trasera en el dispositivo Nota Puede ser que necesite volver a sincronizar su control remoto y el dispositivo WiTouch cuando sustituya la pila en el control remoto Esto garantizará la comunicación entre su control remoto y el dispositivo WiTouch Si su control remoto y el dispo...

Page 10: ...más de cinco días puede indicar un trastorno grave Cese de usar el dispositivo y consulte a su médico El uso del dispositivo teniendo las siguientes condiciones en la piel puede provocar que se agrave la condición de la misma No use el dispositivo sobre o en la cercanía de estas condiciones de la piel piel anormal piel que no esté intacta que esté sucia enferma con heridas abiertas erupciones hinc...

Page 11: ...gel al dispositivo WiTouch Las almohadillas de gel son solamente para el uso de una sola persona Su duración dependerá del tipo de piel grasas y niveles de pH aproximadamente de entre dos a cinco aplicaciones Sustituya las almohadillas de gel una vez que ya no se adhieran por completo Siga estos pasos para aplicar las almohadillas de gel 1 Separe dos almohadillas de gel 2 Retire la cubierta azul d...

Page 12: ...oda estimulación sensorial alta Recomendaciones de tratamiento Puede dejar el dispositivo en posición para tratamientos múltiples durante el día Se apagará automáticamente después de dos 2 horas de inactividad Se recomienda que espere un mínimo de 30 minutos entre tratamientos Efectuando un tratamiento Lea siempre las advertencias de seguridad antes de efectuar un tratamiento Siga estos pasos para...

Page 13: ...completamente cubiertas De artículo Descripción Pila Dispositivo WiTouch El LED del destella una luz amarilla cada dos 2 segundos cuando la unidad está Encendida Pilas alcalinas AAA 2 El LED destella rápidamente en amarillo durante el tratamiento Control Remoto el LED del control remoto destellará en amarillo una vez cuando se presione el botón Inicio paro 1 pila de litio CR2032 Para sustituir las...

Page 14: ... EN NINGÚN CASO POR DAÑOS DIRECTOS O INDIRECTOS Hollywog LLC la Compañía garantiza WiTouch el Producto contra defecto en los materiales y la mano de obra Esta garantía es válida por 1 año 12 meses a partir de la fecha de compra Si estos productos fallan durante el periodo del año de garantía debido a un defecto en los materiales o la maniobra a elección de la empresa la empresa o el distribuidor s...

Page 15: ...IEC 61000 3 2 No se aplica Fluctuaciones de voltaje emisiones de parpadeo IEC 61000 3 3 No se aplica Dirección y la declaración del fabricante inmunidad electromagnética Los WiTouch y WiTouch Pro son intentado para el uso en un ambiente electromagnético especificado abajo El cliente o el usuario del WiTouch o WiTouch Pro deberían asegurar que es usado en tal ambiente Prueba de inmunidad IEC 60601 ...

Page 16: ...propagación electromagnética es afectada por absorción y reflexión de estructuras objetos y personas Dirección y la declaración del fabricante inmunidad electromagnética Los WiTouch y WiTouch Pro son intentado para el uso en un ambiente electromagnético especificado abajo El cliente o el usuario del WiTouch o WiTouch Pro debe usarse en este tipo de ambiente Prueba de inmunidad IEC 60601 Nivel de p...

Reviews: